Magnitude and distribution of occlusal forces on oral implants supporting fixed prostheses: an in vivo study

Authors: Duyck J.1; Van Oosterwyck H.2; Vander Sloten J.2; De Cooman M.3; Puers R.3; Naert I.1

Source: Clinical Oral Implants Research, Volume 11, Number 5, October 2000 , pp. 465-475(11)

Publisher: Blackwell Publishing

Key:
Free Content - Free Content
New Content - New Content
Subscribed Content - Subscribed Content
Free Trial Content - Free Trial Content

Abstract:

Since loading is increasingly believed to be a determining factor in the treatment outcome with oral implants, there is a need to expand the knowledge related to the biomechanics of oral implants. The aim of this study is to gain insight in the distribution and magnitude of occlusal forces on oral implants carrying fixed prostheses. This is done by in vivo quantification and qualification of these forces, which implies that not only the magnitude of the load but also its type (axial force or bending moment) will be registered. A total of 13 patients with an implant supported fixed full prosthesis were selected. Occlusal forces on the supporting implants were quantified and qualified during controlled load application of 50 N on several positions along the occlusal surface of the prostheses and during maximal biting in maximal occlusion by use of strain gauged abutments. The test was conducted when the prostheses were supported by all (5 or 6) implants and was repeated when the prostheses were supported by 4 and by 3 implants only. Despite considerable inter-individual variation, clear differences in implant loading between these test conditions were seen. Loading of the extension parts of the prostheses caused a hinging effect which induced considerable compressive forces on the implants closest to the place of load application and lower compressive or tensile forces on other implants. On average, higher forces were observed with a decreasing number of supporting implants. Bending moments were highest when 3 implants only were used.

Keywords: oral implants; fixed prosthesis; biomechanics; strain gauges; in vivo

Language: English

Document Type: Original article

DOI: 10.1034/j.1600-0501.2000.011005465.x

Affiliations: 1: Department of Prosthetic Dentistry/BIOMAT research group, 2: Department of Mechanics; Division of Biomechanics and Engineering Design, 3: Department of Electrotechnics; ESAT-MICAS, Catholic University Leuven, Belgium *

The full text electronic article is available for purchase. You will be able to download the full text electronic article after payment.

$50.16 plus tax      Refund Policy

 

OR

Back to top

Key:
Free Content - Free Content
New Content - New Content
Subscribed Content - Subscribed Content
Free Trial Content - Free Trial Content
Share this item with others: These icons link to social bookmarking sites where readers can share and discover new web pages.
Page Help Click here for Page Help
Shopping cart
Tools
Sign in






Need to register?
Sign up here
Text size: A | A | A | A